Worksheet: WS-2026-0142
Ready for reviewWorksheet
WS-2026-0142 · EPA 6020B ICP-MS Metals (23-analyte panel)
Source Run
ICP-MS Run 20260616-01 · imported 14:32
Analyst
mvasquez
Verifier
jreeves (pending)
QC Status
All QC within limits (MB clean, LCS 98.2%, MS/MSD 96.4%/95.0%, RPD 1.4%)
Sample results in this worksheet
•••
| Sample | Analytes | QC flags | Reanalysis? | Verifier action | |
|---|---|---|---|---|---|
| MB-0003 | 5 (QC suite) | — | No | Verify | |
| GW-0009 | 23 analytes | — | No | Verify | |
| GW-0010 | 23 (original) + 1 (×10 Zn) | E D reviewer-resolved | Yes — reviewer initiated (×10 dil) | Verify | |
| AF-0005 | 5 (NIOSH 7300 metals) | — | No | Verify | |
| AF-0006 | 5 (NIOSH 7300 metals) | — | No | Verify |
GW-0010 Zn — Reviewer evaluation panel
| Run | DF | Counts (cps) | Within cal? | Result (µg/L) | Flag | Decision |
|---|---|---|---|---|---|---|
| Original (Inj #16, 10:24) | 1 | 64,852,140 | No — above ICAL-5 | >5000 | E | Auto-superseded |
| Reanalysis (Inj #18, 11:24) | 10 | 15,494,896 | Yes | 12,400 | D | Reported |
Both runs were preserved and audited. Reviewer can override the auto-decision (e.g., dispute the dilution result or trigger a third run). Reviewer comments are appended to the audit trail and surface in Stage 2+ packages.