Calibration Summary Layout

Stage 3 adds calibration documentation: all Stage 2A content plus Initial Calibration, Continuing Calibration Verification (CCV), and GC/MS Tune.

CLP-style Calibration Example — Inorganic Data + QC + Calibration

Forms I–XV · Illustrative layout configured to approved lab requirements
CLEARLINE
SDG: CLE-2026-SDG-0142

Form II-IN (Part 1) · Initial Calibration — Zinc (EPA 6020B ICP-MS)

StandardTrue (µg/L)Response (cps)Predicted%Diff
ICAL-1 (blank)0.01,2400.0
ICAL-25.062,4505.020.4%
ICAL-350.0624,15050.10.2%
ICAL-45006,241,500500.40.08%
ICAL-5 (high)5,00062,485,2005,0010.02%
Curve fit: Linear, no weighting · r² = 0.99998 (criterion: ≥0.998) ✓  Slope: 12,496.4 · Intercept: 1,240 cps · Date: 2026-06-16 07:14 · Analyst: mvasquez

Form II-IN (Part 2) · Continuing Calibration Verification (CCV)

CCV #TimeAnalyteTrueFound%RLimitsPass
CCV-109:42Zinc10098.298.2%90–110%
CCV-109:42Lead100101.4101.4%90–110%
CCV-211:18Zinc10099.699.6%90–110%
CCV-211:18Lead100100.1100.1%90–110%
CCV-313:55Zinc10097.997.9%90–110%

Form II-IN (Part 3) · Initial Calibration Blank (ICB) & Continuing Calibration Blank (CCB)

Blank TypeTimeZn (µg/L)Pb (µg/L)Cd (µg/L)CriterionPass
ICB07:24< 0.5< 0.05< 0.05< CRDL/2
CCB-109:44< 0.5< 0.05< 0.05< CRDL/2
CCB-211:20< 0.5< 0.05< 0.05< CRDL/2

Form XV-IN · Instrument Tune (ICP-MS)

MassElementResolution (amu)%RSD (10 readings)CriterionPass
9Be0.752.1%< 5%
59Co0.781.4%< 5%
89Y0.761.8%< 5%
205Tl0.792.6%< 5%
Instrument: Agilent 7900 ICP-MS · Tune date: 2026-06-16 06:42 · Operator: mvasquez · Tune file: ZN_CAL_20260616.tune

Interference Check Sample (ICS-A / ICS-AB)

SampleAnalyteTrueFound%RLimitsPass
ICS-AZinc50049599.0%80–120%
ICS-ABZinc500504100.8%80–120%